Display
Operating
temperature
Storage
temperature
Ingress Protection
Humidity
Shock/Vibration
Pollution Degree
EMC
Electrical safety
Pressure safety
Approved
Battery power
06 June 2013
LCD: Color display with touch-screen
-10 to 50°C (14 to 122°F)
-20 to 70°C (-4 to 158°F)
IP55 (Druck DPI 620 Genii calibrator only)
0 to 90% relative humidity (RH) non-condensing
MIL-PRF-28800F for class 2 equipment
2
Electromagnetic compatibility: BS EN 61326-
1:2006
Electrical - BS EN 61010:2010
Pressure Equipment Directive - Class: Sound
Engineering Practice
(SEP)
CE Marked
Lithium-Polymer battery (GE Part number: 191-
356)
Capacity: 5040 mAh (minimum), 5280 mAh
(typical), Nominal voltage: 3.7 V.
Charge temperature: 0 to 40°C (32 to 104°F)
outside this range, charging stops.
Discharge temperature: -10 to 50°C
Druck DPI620 Genii