Assembly Instructions; Pressure Connections; Procedure (Attaching External Equipment); Measure Pressure - GE Measurement & Control Druck DPI 620 Genii User Manual

Advanced modular calibrator
Hide thumbs Also See for Measurement & Control Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
3.1

Assembly Instructions

3.1.1 Pressure Connections

Pressurized gases and fluids are dangerous. Before
attaching or disconnecting pressure equipment, safely
release all the pressure.
The pressure ports for external equipment use "Quick fit"
pressure adaptors
3.2

Procedure (Attaching External Equipment)

1.
Remove the adaptor from the pressure port
2.
Use an applicable seal for the pressure connection
i
ii
iii
3.
Attach the adaptor to the external equipment if
necessary use an alternative adaptor
4.
Tighten to the applicable torque
5.
Attach the adaptor to the MC 620 carrier and hand-
tighten
3.3

Measure Pressure

When the pressure indicator assembly is complete use the
06 June 2013
WARNING
NPT type: Use an applicable sealant on the
thread
BSP (parallel) type: Use the applicable bonded
seal at the bottom
BSP (parallel) type, 100 bar (1500 psi) or less: a
bonded seal at the top is permitted
Druck DPI620 Genii
Page 67 of 179

Advertisement

Table of Contents
loading

Table of Contents