8.3

Procedures (CH1/CH2): Current (Source)

Do the procedure as follows:
1. Connect the applicable calibration equipment
• CH1/CH2 (24 mA range): (Ref: Section 2.4 and Section
2.9)
• CH2 (-24 mA range only): ): (Ref: Section 2.4 and
Section 2.9)
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on)
3. Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration (0.2 mA and FS):
• CH1 (one range): 24 mA
• CH2 (two ranges): 24 mA (reverse) and 24 mA (forward)
Note: Input positive values for the forward and reverse
calibration
06 June 2013
(Ref: Table 8-1):