User Manual
• Check the error is within limits (Ref:
• Table
Table 8-8 Hz error limits (measure/source)
Measure/
source Hz
100
990
Table 8-9 kHz error limits (measure/source)
Measure/
source kHz
10.0000
50.0000
06 June 2013
or Table 8-9)
8-8
Calibrator
uncertainty
(Hz)
0.0002
0.0005
Calibrator
uncertainty
(kHz)
0.00002
0.00002
Permitted
DPI 620 error (Hz)
(measure)
0.0023
0.0050
Permitted
DPI 620 error
(kHz)
(measure)
0.00023
0.00035
Druck DPI620 Genii
(source)
0.0026
0.0053
(source)
0.000067
0.000185
Page 147 of 179