5. Check the calibration is correct:
• Select the applicable pressure function
• Apply the following pressure values (absolute sensors)
• 0, 20, 40, 60, 80, 100 (%FS)
• Go back to 0 in the same steps
• Apply the following pressure values (gage sensors
• 0, 20, 40, 60, 80, 100 (%FS)
• Go back to 0 in the same steps
6. Check the error is within limits (Ref: Table Gage Ranges
and Absolute Ranges)
• Gage ranges or absolute ranges tables (Ref:
datasheet
• Use the values in the total uncertainty column
• The specified values include an allowance for
temperature changes, reading stability for one year,
and the uncertainty of the standard used for
calibration