Advertisement

User Manual
5.4.1 Sequence to Upload and Download File
Step
1. Connect Standard USB lead (as supplied) to the Druck.
DPI 620 Genii calibrator.
2. Connect the lead to the USB port on the computer that
has the calibrator manager installed.
3. Use 4 Sight to set-up the procedure and create a work
order for the device.
The procedure includes the parameters for the
calibration, the number of test points, the relationship
and pass/fail tolerance.
4. Use the Download button in calibrator manager to
Download the file to the Druck DPI 620 Genii calibrator.
A communications symbol will be displayed at the
bottom of the screen.
5. Tap the Run Procedure option.
6. In the Results window, select the filename specified in
calibrator manager.
7. Enter the User ID and the DUT Serial Number.
8. Tap on the Start button. The Procedure sets up the
necessary Channel options e.g. mA and Volts.
9. Use the Take Reading button at each point specified by
the procedure. A prompt shows for each point.
10. When all the readings are complete, tap on the Exit
button. Look at the results on the display (As found/As
Left).
11. To complete the process, use the calibrator manager to
Upload the file back into the 4 Sight database
06 June 2013
Procedure
Druck DPI620 Genii
Page 88 of 179