GE Measurement & Control Druck DPI 620 Genii User Manual page 141

Advanced modular calibrator
Hide thumbs Also See for Measurement & Control Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
Table 8-4 Millivolts (measure) error limits
Applied
Table 8-5 Voltage (measure) error limits
Applied
06 June 2013
mV
uncertainty
±2000
±1000
±200
±100
0 (short
circuit)
V
uncertainty
±30
±21
±20
±10
±5
0
Calibrator
(mV)
0.051
0.040
0.0051
0.0040
0.0036
Calibrator
(V)
0.00052
0.0004
0.00031
0.00016
0.00008
0.000024
Druck DPI620 Genii
Permitted
DPI 620 error
(mV)
0.14
0.1
0.017
0.0125
0.008
Permitted
DPI 620 error
(V)
0.0021
0.0018
0.0009
0.00065
0.00053
0.0004
Page 141 of 179

Advertisement

Table of Contents
loading

Table of Contents