Procedures (Ch1): Dc Mv/Volts (Source) - GE Measurement & Control Druck DPI 620 Genii User Manual

Advanced modular calibrator
Hide thumbs Also See for Measurement & Control Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
8.5

Procedures (CH1): DC mV/Volts (source)

Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref:
Table 8-1)
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on)
3. Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration (Zero and +FS) for the applicable range:
mV (source) ranges
2000 mV
4. Check the calibration is correct:
• Select the applicable Millivolts or Voltage (measure)
function (ref Section 0). Apply the input values that are
applicable to the calibration:
mV: 0, 100, 200, 1000, 2000
Volts (V): 0, 5, 10, 15, 20.
• Check the error is within limits (Ref: Table 8-6 or Table
8-7)
06 June 2013
Druck DPI620 Genii
Volts (source) ranges
20 V
Page 142 of 179