GE Measurement & Control Druck DPI 620 Genii User Manual page 135

Advanced modular calibrator
Hide thumbs Also See for Measurement & Control Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
Function
AC mV
(CH1)
AC Volts
(CH1)
Pressure (P1 or P2)
IDOS
Before starting the calibration, check the time and date on the
instrument are correct
06 June 2013
Calibration equipment
(ppm = parts per million)
Thermocouple temperature reference
unit (0°C)
Accuracy: 30 mK
AC mV calibrator.
Accuracy - AC mV measure,
Refer to Table 8-15
AC Volts calibrator.
Accuracy - AC Volts measure,
Refer to Table 8-16
Genii Module Carrier MC620G or Genii
Pressure Base
PV62XG:
Range 25 mbar/0.36 psi: total
uncertainty of 0.015%
reading or better
Ranges > 25 mbar/0.36 psi: total
uncertainty of 0.01%
reading or better
UPM only. Refer to the user manual for
the IDOS UPM.
Druck DPI620 Genii
Page 135 of 179