1.6

Electrical warnings

• The DC input to the DPI620 Genii is rated at 5V (+/-5%) 2
Amps
• To prevent electrical shocks or damage to the instrument,
do not connect more than 30V CAT I between the
terminals, or between the terminals and the ground
(earth)
• External circuits should have appropriate insulation to the
mains
• To prevent electrical shocks, use only the GE specified
AC probe (Part: IO620-AC) to measure AC voltages that
are more than 20 Vrms. Do not connect more than
300V CAT II between the IO620-AC leads, or between
the leads and the ground (earth). Attach it to the
specified connections only
• This instrument uses a Lithium-Polymer (Li-Polymer)
battery pack. To prevent an explosion or fire, do not
short circuit, do not disassemble, and keep it safe
from damage
• To prevent an explosion or fire, use only the GE
specified battery (Part: 191-356), power supply (Part:
191-339) and battery charger (Part: IO620-CHARGER)
• To prevent battery leakage or heat generation, only
use the battery charger and power supply in the
temperature range 0°C to 40°C (32°C to 104°F
06 June 2013
Druck DPI620 Genii