Active Power Accuracy - Teledyne Everywhereyoulook T3PM1100 User Manual

Digital power meter
Hide thumbs Also See for Everywhereyoulook T3PM1100:
Table of Contents

Advertisement

changes after zero-level
compensation or range
change
Accuracy when the crest
factor is set to 6 or 6A
Accuracy changes caused
by data update interval

Active Power Accuracy

Item
Requirements
Accuracy
Influence of power factor
202
5 mA/10 mA/20 mA/50 mA/100
mA/200 mA ranges
0.5 A/1 A/2 A/5 A/10 A/20 A
ranges
External current sensor input (/EX1) 1 mV/°C
External current sensor input (/EX2) 50 µV/°C
accuracy obtained by doubling the measurement
range error for the accuracy when the crest factor is
set to 3
When the data update interval is 100 ms, and Auto,
add 0.05% of reading to the 0.1 Hz to 1 kHz
accuracy.
Specifications
Same as the conditions for voltage and current.
Power factor
DC
0.1Hz ≤ f < 45 Hz
45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1kHz
1 kHz < f ≤ 10 kHz
10 kHz < f ≤ 100
kHz
When power factor (λ) = 0 (S: apparent power)
± 0.1 % of S for 45 Hz ≤ f ≤ 66 Hz
± {(0.1 + 0.15 × f) % of S } for up to 100 kHz as
reference data
•f is frequency of input signal in kHz
When 0 < λ < 1 (Φ: phase angle of the Voltage and
current)
T3PM1100 User Manual
5 µA/°C
500 µA/°C
1
(0.1 % of reading + 0.2 % of
range)
± (0.3 % of reading + 0.2 % of
range)
± (0.1 % of reading + 0.05 %
of range)
± (0.2 % of reading + 0.2 % of
range)
± (0.1 % of reading + 0.3 % of
range) ± [{0.067x(f-1)}% of
reading]
± (0.5 % of reading + 0.5 % of
range) ± [{0.09x(f-10)}% of
reading]

Advertisement