Parameter
Return parameter 0
Example

[:INPut]:POVer

Description
Query Syntax
Return parameter Bit0
Example

[:INPut]:CRANge

Description
Query Syntax
Return parameter Bit0
156
<Boolean>0 OFF
<Boolean>1 ON
Turn the frequency filter function off.
1
Turn the frequency filter function on.
:INPUT:FILTER:FREQUECNY OFF
:INPUT:FILTER:FREQUECNY?
->:INPUT:FILTER:FREQUECNY 0
Returns the peak over-range information.
[:INPut]:POVer?
Voltage peak over-range is occurring.
Bit1
Current peak over-range is occurring.
:INPUT:POVER?
->:INPUT:POVER 1
Sets or returns the check range status.
[:INPut]:CRANge?
The voltage is at the condition for
reducing the auto range or less.
Bit1
The voltage exceeds the condition for
raising the auto range.
Bit2
The voltage is over-range.
Bit3
The voltage is peak over-range.
Bit4
The current is at the condition for
reducing the auto range or less.