3.5 Testing the System

Before plugging the instrument into the power source:
 Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.
 Check the integrity and accuracy of the electrical connections.
Make sure there are no exposed conductors.
 Check that the pressure and flow of all gases are within the
recommended levels, and appropriate for your application.
Power up the system, and test it by performing the following
operations:
1. Repeat the Self-Diagnostic Test as described in chapter 4,
section 4.3.5.

3.6 Warm Up at Power Up

Every time the unit is turned on, the instrument stays with the
introduction screen for thirty minutes. This is to allow the cell to come
up to temperature (60 ° C). The only way to bypass this warm up period
is by pressing skip button on the touchscreen.
The instrument warms up for half an hour so that it will not receive
a remote calibration signal, send false readings to a monitor system, or
be calibrated by an untrained operator while the cell is cold.
NOTE: There is temperature readout on the display while on
warm up mode and normal operation mode to monitor if
TCD is at correct temperature.