[:INPut]:CFACtor
Description
Syntax
Query Syntax
Parameter/
Return parameter
Example
[:INPut]:WIRing
Description
Syntax
Query Syntax
Parameter/
Return parameter
Example
[:INPut]:MODE
Description
Syntax
146
Sets or returns the crest factor.
[:INPut]:CFACtor {3|6|A6}
[:INPut]:CFACtor?
3
crest factor 3.
6
crest factor 6.
A6
Display range expand mode (6A) for
crest
factor 6.
:INPUT:CFACTOR 3
:INPUT:CFACTOR?
->:INPUT:CFACTOR 3
Sets or returns the wiring system.
[:INPut]:WIRing {P1W2 }
[:INPut]:WIRing?
P1W2
Single-phase, two-wire system. (For
the T3PM1100, the wiring system is
fixed to P1W2)
:INPUT:WIRING P1W2
:INPUT:WIRING?
->:INPUT:WIRING P1W2
Sets or returns the voltage and current
measurement mode.
[:INPut]:MODE {DC|AC/RMS|ACDC| VMEan}
T3PM1100 User Manual
Set
Query
Set
Query
Set
Query