Status:filter<X - Teledyne Everywhereyoulook T3PM1100 User Manual

Digital power meter
Hide thumbs Also See for Everywhereyoulook T3PM1100:
Table of Contents

Advertisement

Note
:STATus:FILTer<x>
Description
Syntax
Query Syntax
Parameter/
Return parameter
Example
• If no errors have occurred, 0,"No error" is returned.
• User can use the :STATus:QMESsage command to
specify whether the message is included.
• Error message description:
Error_103: Invalid separator
Error_104: Data type error.
Error_108: Parameter not allowed.
Error_109: Missing parameter.
Error_113: Undefined header.
Error_131: Invalid suffix.
Error_141: Invalid character data.
Error_221: Setting conflict.
Error_222: Data out of range.
Error_813: Invalid operation.
Sets or returns the transition filter.
:STATus:FILTer<x> {RISE|FALL|BOTH|NEVer}
:STATus:FILTer<x>?
<x>
1~16
RISE
An event is set when the bit changes from 0
to 1.
FALL
An event is set when the bit changes from 1
to 0.
BOTH
An event is set when the bit changes either
from 1 to 0 or form 0 to 1.
NEVer
An event is never trigger.
:STATUS:FILTER2 RISE
:STATUS:FILTER2?
-> :STATUS:FILTER2 RISE
COMMAND OVERVIEW
Set
Query
189

Advertisement

Table of Contents
loading

Table of Contents