To Test Bandwidth; Equivalent Time Test - Agilent Technologies 54845B Manual

Table of Contents

Advertisement

To test bandwidth

This test checks the bandwidth of the oscilloscope. The Agilent 54845B bandwidth at 1.5 GHz
and the Agilent 54846B bandwidth at 2.25 GHz oscilloscopes are checked using aliasing since
the trigger specification of these oscilloscopes is 1 GHz. The 1.5 GHz or the 2.25 GHz signal is
viewed with a time/division setting of approximately 50 µs/div where the on-screen display is an
untriggered sinewave that appears to be at a lower frequency, however the vertical response of
the amplifiers is correct.
Specification
Equivalent Time:
50Ω: 2.25 GHz (Agilent 54846B), 1.5 GHz (Agilent 54845B), 1 MΩ: 500 MHz (with Agilent
1161A probe)
Real Time:
50Ω: for Agilent 54846B — 2.25 GHz (2-channel mode), 1.0 GHz (4-channel mode)
50Ω: for Agilent 54845B — 1.5 GHz (2-channel mode), 1.0 GHz (4-channel mode)
1 MΩ: 500 MHz (with Agilent 1161A probe)
Equipment Required
Equipment
Critical Specifications
10—2250 MHz at ≈200 mVrms
Signal Generator
1—1.5 GHz ±3% accuracy
Power Meter/Sensor
Power Splitter
outputs differ by <0.15 dB
Cable
Type N (m) 24 inch
Termination
50Ω, BNC Connectors
Adapter
Type N (m) to BNC (m)
Probe
No Substitute
Probe Tip Adapter
1160 Series to BNC

Equivalent Time Test

Connect the equipment.
1
a With the N cable, connect the signal generator to the power splitter input. Connect the
power sensor to one output of the power splitter.
b With an N-to-BNC adapter, connect the other splitter output to the channel 1 input.
Press Default Setup to set the oscilloscope to default conditions.
2
Select Acquisition from the Setup menu. Select Equivalent Time sampling mode. Click
3
Close. Select Display from the Setup Menu. Unselect Connect Dots mode. Click Close.
Set the vertical scale for Channel 1 to 100 mV/div using the knob. Select 50Ω input
4
impedance.
Set the sweep speed to 50 ns/div using the horizontal sweep speed knob.
5
Set the signal generator for 10 MHz at +5.0 dBm.
6
The signal on the oscilloscope screen should be about five cycles at six divisions amplitude.
Bandwidth Check Limits
Do not exceed 6 divisions when making the bandwidth check.
Chapter 3: Testing Performance

To test bandwidth

Recommended Model/Part
Agilent 8664A
Agilent EPM-441A/8482A
Agilent 11667A
Agilent 11500B
Agilent 10100C
Agilent 1250-0082
Agilent 1161A
Agilent 5063-2143
3–13

Advertisement

Table of Contents
loading

This manual is also suitable for:

54846b

Table of Contents