15 - 2
Checking the compression device and OPCOMP
6.
7.
1
Fig. 3
Display at the lower part of the stand
8.
NOTICE

Decompression button (only with separate control console)

Check the decompression button
i.e. that the compression plate moves upwards when the button is pressed.
Mammomat 1000/3000
Installation and Start-Up
Press the foot switch for motorized decompression to move the compression plate
upwards while at the same time pressing the compression plate in the opposite
direction by hand.
The upward movement of the compression plate shall now stop (safety switch).
Set the compression to 7 kg by using the presetting knob (5/Fig. 1).
The preset compression force appears on the display at the lower part of the stand
(1/Fig. 3).
2
Press the foot switch for motorized compression to move the compression plate
downwards while simultaneously pressing the compression plate in the opposite
direction by hand.
The compression plate shall now slow down and stop, when the preset
compression force is reached.
During compression, the motorized vertical adjustment as well as
the rotary motion of the swivel-arm system shall be blocked.