MAMMOMAT 1000/3000
Installation
Installation and Start-Up Instructions
Register 3
Print No.: SPB7-230.033.07.03.02
SPB7-230.033.07.02.02
Replaces:
© Sieme ns AB 1999
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 11.99
65 27 464

Advertisement

Table of Contents
loading

Summary of Contents for Siemens MOMMOMAT 1000

  • Page 1 MAMMOMAT 1000/3000 Installation Installation and Start-Up Instructions © Sieme ns AB 1999 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_...
  • Page 2 0 - 2 Revision Chapter Page Revision Mammomat 1000/3000 Register 3 SPB7-230.033.07 Page 2 of 2 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 3: Table Of Contents

    EMC measures at the cable entry ......5 - 1 Siemens-Elema AB Register 03 SPB7-230.033.07...
  • Page 4 Checking X-ray tube high voltage, tube currrent and mAs values ...10 - 2 MAMMOMAT 1000/3000 Register 03 SPB7-230.033.07 Page 2 of 4 Siemens-Elema AB Installation and Start-up Rev. 03 11.99 SPS-UD...
  • Page 5 16 ______Checking the exposure blocking _________________________________ 16 - 1 Checking the exposure blocking ......16 - 1 Siemens-Elema AB Register 03 SPB7-230.033.07...
  • Page 6 Measuring protocol ........20 - 3 MAMMOMAT 1000/3000 Register 03 SPB7-230.033.07 Page 4 of 4 Siemens-Elema AB Installation and Start-up Rev. 03 11.99 SPS-UD...
  • Page 7: General

    Fixed object-table arm ("Wing 2" no counterweight) Generator with integrated Separate generator control panel and radiation shield Separate control console Free-standing radiation shield Fig. 1 Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 6 Mammomat 1000/3000 Solna, Sweden Rev. 03 11.99...
  • Page 8: Mammomat 3000

    Depending on the installation configuration, some points may be omitted. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 6 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 9: Meters And Appliances, Tools

    • Densitometer e.g PDA 81. • Service PC (e.g Siemens-Nixdorf PCD 3NSX/20 or similar) with connecting cable (PC- Generator), part no. 99 00 440 RE999. • Normi 7 test body (provided by the customer) or SIB phantom Type 42 001 (PTW or INAK).
  • Page 10: Important Notes On Start-Up

    Furthermore, the front page of the acceptance test certificate must also be completed with the operator´s data. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 4 of 6 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 11: Checking And Recording For The Dhhs Area Of Application

    The MAMMOMAT has been tested and adjusted at the factory, and should therefore be ready for operation after completion of the installation. The equipment is set to 400 V, 2-pase, on delivery. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 5 of 6...
  • Page 12 1 - 6 Prerequisites This page intentionally left blank. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 6 of 6 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 13: Protective Measures

    Protective Measures after removing the covers from the the stand. Remove the protective strips only when the covers are to be mounted or when vertical adjustment of the swivel-arm system is necessary. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 2...
  • Page 14 2 - 2 Protective measures This page intentionally left blank. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 2 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 15: General

    As a general rule, the directional marks on the crates should be observed during trans- port, storage and unpacking. The stand crate is bolted together, while the cardboard box is secured with plastic straps. WARNING Wear safety footwear! Siemens-Elema AB Register 03 SP B7-230.033.07 Page 1 of 8 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 16: Unpacking The Stand

    Upend the stand (4), with pallet (5) (two persons are required). Loosen the four bolts (6) and remove the pallet (5). Mammomat 1000/3000 Register 03 SP B7-230.033.07 Page 2 of 8 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 17: Removing The Transport Safeguards

    Fig. 3 Transport safeguard for the lifting carriage Loosen the two screws (1/Fig. 3) and remove the red transport safeguard (2/Fig. 3). Reinsert screws (1/Fig. 3), they might be needed for service purposes. Siemens-Elema AB Register 03 SP B7-230.033.07 Page 3 of 8...
  • Page 18: Transport Safeguard For The Rotary Motion

    Screw and spacer inserted from the inside of the stand Tighten the screw. Make sure that the transport safeguard is parallell to the stand. Mammomat 1000/3000 Register 03 SP B7-230.033.07 Page 4 of 8 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 19: Protective Strips For The Metal Curtain

    Holders (2) for storing the protective strips when not in use, are provided on both sides of the curtain, see Fig. 7. Siemens-Elema AB Register 03 SP B7-230.033.07 Page 5 of 8...
  • Page 20: Unpacking The Generator And Mounting The Radiation Shield (Optional)

    Remove the protective cover from the generator. Upend the generator (1/Fig. 8) and lift it off the pallet (2/Fig. 8) (two persons are required). Mammomat 1000/3000 Register 03 SP B7-230.033.07 Page 6 of 8 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 21 Mount the stanchions and lead-glass unit with 6 screws onto the generator (3, 4 / Fig. 8) (two persons are required). Cover the holes with plastic caps (7/Fig. 9). Fig. 9 Mounting the radiation shield Fig. 10 Siemens-Elema AB Register 03 SP B7-230.033.07 Page 7 of 8 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 22 3 - 8 Preparatory work This page intentionally left blank. Mammomat 1000/3000 Register 03 SP B7-230.033.07 Page 8 of 8 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 23: Arranging The Components

    (option). The connection cable to the generator measures 10 m. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 6...
  • Page 24: Free-Standing Radiation Shield (Option)

    EMC (Electro Magnetic Compatibility) requirements. Remove the front cover (3) from the generator. Fig. 2 Removing the generator cover Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 6 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 25: Installing The Cable Ducts

    At the generator, cover the cable entry not used with the enclosed cover plate (7). Installations with separate control console For the connection cable separate control console – generator, additional cable ducts are required, unless the cables are to be laid underfloor. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 3 of 6...
  • Page 26: Laying The Cable Harness

    Anode rotation cable X9 Filament cable X8 Mains supply cables L1, L2 Protective ground Decompression-button cable (only with separate control console) Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 4 of 6 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 27: Installations With Separate Control Console

    ∅ 6.5 mm Plastic cap Wall-mounted cable duct is recommended ∅ 6.5 mm Vertical mounting Horizontal mounting Fig. 5 Wall-mounted control console Siemens-Elema AB Register 3 SP B7-230.033.07 Page 5 of 6 Mammomat 1000/3000 Solna, Sweden Rev. 03 11.99 SPS-UD...
  • Page 28: Aligning The Stand

    If bolting to the floor is required, please refer to Chapter 19 "Bolt- ing the stand/generator to the floor" on page 19 - 1. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 6 of 6 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 29: Emc Measures

    Lay the cable harness from the stand to the generator and leave approximately 3 cm of the shielding at the U-bracket, see Fig. 1. Stow away excessive cable lengths as shown in Fig. 2. Siemens-Elema AB Register 3 SP B7-230.033.07...
  • Page 30: Note On Separate Control Console And Separate Generator

    5 - 8 for information on how to install the additional connection cable between the control console and the generator. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 14 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 31: Emc Measures On The Bottom Plate

    Fig. 7 in this chapter. NOTICE Do not remove insulation and do not ground the shields of the IONTOMAT signal cables to X10 and X11. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 3 of 14 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 32: Connecting The Stand Cable-Harness

    Make sure that there aren’t any air bubbles between the disc and the connector. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 4 of 14 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 33: Connecting The High-Voltage Connector

    Plug in the high-voltage connector (1) at H1. Tighten the sleeve nut (2) by hand. Position the high-voltage cable (3) to the right in the generator and fasten with cable ties (4). Siemens-Elema AB Register 3 SP B7-230.033.07 Page 5 of 14...
  • Page 34: Connecting The Cable Harness To The Generator

    X10 and X11 at the end of the cable. Connect the IONTOMAT signal cable connectors, X10 and X11, to D700 X10 (5) and X11 (6). Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 6 of 14 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 35: Rotating Anode Cable X9

    Connect the two blue free wires from connector D711 X9 (1) to capacitor C51 (2). Clamp the cable to the chassis with a cable clamp (3). CAUTION When fitting the cable clamp, ensure satisfactory contact between shield braid (4) and chassis. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 7 of 14 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 36: Installations With Separate Control Console And Separate Generator

    U-bracket and secure with a hose clamp as shown in Fig. 9. Fig. 9 Shielding the control console cable Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 8 of 14 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 37 T3/X30 at the side of the circuit board chassis. Connect the ground wire to the protective ground terminal below fuse F10, see Fig. 11. Protective ground T3/X30 Fig. 11 Connecting the cables from the separate control console Siemens-Elema AB Register 3 SP B7-230.033.07 Page 9 of 14 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 38: Main Voltage Connection

    Labels are included in the installation material. For more details about changing mains, insertion or removal of NOTICE jumpers, fuses, labeling etc, see Wiring Diagram. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 10 of 14 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 39: Connecting The Incoming Mains To The Stand (400 V, 2-Phase)

    GENERATOR STAND Incoming mains 208, 230, 240, 277 or 400 VAC P.E N L1 L2 L3 X899 X881 Fig. 14 Incoming mains, 2-phase Siemens-Elema AB Register 3 SP B7-230.033.07 Page 11 of 14 Mammomat 1000/3000 Solna, Sweden Rev. 03 11.99...
  • Page 40: Connecting The Mains Supply To The Generator (400 V, 2-Phase)

    Wire marked L1 to F10 and L2/N to F20. Connect the protective ground wire from the stand to the generator (3). Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 12 of 14 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 41: Measures For Changing From 2-Phase To 1-Phase Connection

    Wiring Diagram for details. Connect the mains supply cable between stand and generator to fuses F10 and F20 on the generator. Wire marked L1 to F10 and L2/N to F20. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 13 of 14...
  • Page 42 5 - 14 Cable connections This page intentionally left blank. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 14 of 14 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 43: Checks Before Powering Up The Generator

    Remove fuses F10 (1) and F20 (2) or, when single-phase supply, fuse F10 (1). Connect line resistance meter to fuse holders F10 and F20. Mains supply "ON". Carry out measurement. Mains supply "OFF". Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 4 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 44: Checking The Line Voltage In The Generator

    Measure the voltage at the output of line filter Z1 it must lie within 195 V- 264 V. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 4 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 45: Checking The Supply Voltages

    System "ON". Check that the following orange diodes are "ON" on p.c. board D704: V39 (+15 V) V38 (-15 V) V41 (+5 V) V40 (+24 V) Siemens-Elema AB Register 3 SP B7-230.033.07 Page 3 of 4 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 46 6 - 4 Mains connection and power supply This page intentionally left blank. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 4 of 4 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 47: Attaching The Swivel-Arm Covers

    Before installing the side covers, the cables are to be connected. MAMMOMAT 1000/3000 Connect cables X807 (1) to control-button circuit boards D807 (2). Connect ground wires (3) to patient handles (4). Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 2...
  • Page 48: Attaching The Side Covers

    The front cover must not press against the collimator (risk of damage). Fasten the front cover to the side covers with two short screws (5). Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 2 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 49: Microprocessors

    P.C. board D711: • V9 "ON" (line voltage on). • V13 "OFF" (system stand-by). • V11 "ON" (system on). • V16 "ON" (intermediate circuit - filament). Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 2 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 50 8 - 2 Checking the microprocessors This page intentionally left blank. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 2 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 51: Checks Without High Voltage

    Press the exposure buttons on the control panel. The rotating anode shall start up. At the end of the limit time, the rotating anode is braked. Release the exposure buttons. System "OFF". Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 2...
  • Page 52 9 - 2 Checks without high voltage This page intentionally left blank. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 2 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 53: Preparation

    Set switch S2 (SS) to "ON" (upper position) on P.C. board D702. Remove the jumper and connect the mAs meter to mAs measuring sockets X3 and X4 on P.C. board D710. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 6...
  • Page 54: Checking X-Ray Tube High Voltage, Tube Currrent And Mas Values

    Chapter 12 "Further programming". The tube current reduction is effective only in IONTOMAT mode with Bucky. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 6 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 55: Oscilloscope Diagrams

    127 ms 20 ms/T 160 ms 20 ms/T NOTE Record the measured kV and mAs value in the measuring proto- col, page 20 - 3. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 3 of 6 Mammomat 1000/3000 Solna, Sweden Rev. 03 11.99...
  • Page 56 2 V/T (1 T <=> 80 mA) 150 mA 2 V/T (1 T <=> 80 mA) 0.53 s 0.1 s/T 0.67 s 0.1 s/T Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 4 of 6 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 57 Small focus is selected by attaching a magnification table. NOTICE NOTICE Record the measured kV and mAs value in the measuring proto- col, page 20 - 3. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 5 of 6 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 58 1 V/T (1 T <=> 40 A) 100 ms/T) 100 ms/T ) total exposure time dependent on programmed sensitivity Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 6 of 6 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 59: Checking And Programming With The Service Pc

    Valid entries from 0 to 31 in 1/2 E.P. Input of decimal point not required! 1 Help 2 Save 10 Quit Fig. 1 Values from 0 to 31 can be entered. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 12 Mammomat 1000/3000 Solna, Sweden Rev. 03 11.99...
  • Page 60: Sensitivity Correction

    IONTOMAT PM - Correction curve Curve number Speed H Speed D 1 Help 2 Save 10 Quit Fig. 3 Values shown as delivered. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 12 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 61: Other Film/Screen Combinations

    Test the chosen curve/curves with the other anode/filter combinations as well. NOTICE The film density levels may vary from cassette to cassette. Therefore, be sure to use one and the same cassette for all test exposures. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 3 of 12 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 62: Correction Curve Diagrams

    25 kV Mo/Mo Grid thin object quotient thick object Diagram 1 25 kV Mo/Mo Mag thin object quotient thick object Diagram 2 Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 4 of 12 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 63 27 kV Mo/Mo Grid thin object quotient thick object Diagram 3 27 kV Mo/Mo Mag. thin object quotient thick object Diagram 4 Siemens-Elema AB Register 3 SP B7-230.033.07 Page 5 of 12 Mammomat 1000/3000 Solna, Sweden Rev. 03 11.99 SPS-UD Installation and Start-Up...
  • Page 64 30 kV Mo/Mo Grid thin object quotient thick object Diagram 5 30 kV Mo/Mo Mag. thin object quotient thick object Diagram 6 Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 6 of 12 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 65 35 kV Mo/Mo Grid thin object quotient thick object Diagram 7 35 kV Mo/Mo Mag. thin object quotient thick object Diagram 8 Siemens-Elema AB Register 3 SP B7-230.033.07 Page 7 of 12 Mammomat 1000/3000 Solna, Sweden Rev. 03 11.99 SPS-UD Installation and Start-Up...
  • Page 66: Checking The Film Density

    Repeat this procedure on the other "wing" (base plate). The final values are dependent on the type of cassette used and desired optical density. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 8 of 12 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 67: Checking The Automatic Transparency Adaption

    These limits apply for film with max. gamma 4.0 measured at the density chosen. NOTICE The film density levels may vary from cassette to cassette. There- fore, be sure to use one and the same cassette for all test expo- sures. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 9 of 12 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 68: Checking The Automatic Transparency Adaption And Opdose (Mammomat 3000 Only)

    1.5 O.D. (or the density preferred by the customer). Store with the memory button. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 10 of 12 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 69: Checking Cut-Off Dose And Resolution (§16 Germany)

    G(2) Calculate and record the cut-off dose as follows: G 1 ( ) ⋅ ------------- - B 2 ( ) g 2 ( ) Siemens-Elema AB Register 3 SP B7-230.033.07 Page 11 of 12 Mammomat 1000/3000 Solna, Sweden Rev. 03 11.99...
  • Page 70: Output Values Of Constancy Testing (§16 Germany)

    (steel balls). Read off the pen dosimeter and record the dose value. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 12 of 12 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 71: Setting The Real Time Clock

    Tube-current reduction (TCR) at short exposure times. TCR switch is ON/OFF. Exposure time with 50 mA tube current is 100 ms. Switch ON or OFF with <↓>. Save the programmed values with <F2>. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 2 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 72: Reducing The Generator Power

    Type in line impedance ____ Ω. Enter the measured line impedance. Calculated power is ___ kW. Max. possible tube power is displayed. Save the programmed values with <F2>. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 2 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 73: Rotary Motion

    (+135 degrees clockwise and -180 degrees counterclockwise). The projection angle is shown on the display at the lower part of the stand. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 1 of 4 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 74: Vertical Adjustment

    All eight vertical adjustment buttons (four on either side) must be checked. The switching off in upper and lower end position is effected by switches S881 and S884 respectively. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 4 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 75: Emergency Stop

    The button shall latch and all motorized movements of the equipment shall be blocked ("Er 824" shall appear). To reset the emergency-stop button, turn it clockwise. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 3 of 4 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 76 13 - 4 Checking the swivel-arm system This page intentionally left blank. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 4 of 4 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 77: Checking And Adjusting The Field Light Time

    Switch on the field light by pressing switch (1/Fig. 1) or by pressing the compression foot switch (2/Fig. 2). The field light shall go out automatically after a preset period of time (normally 20 s.). Fig. 2 Foot switch Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 2 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 78: Adjusting The Field Light Time

    Select Miscellaneous <↑> <↓> <ENTER>. Select Illumination time <↑> <↓> <ENTER>. Type in illumination time and save with < F2>. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 2 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 79: Checking The Compression Device

    Place a piece of soft material on the object table, for example a rolled up towel or a piece of foam-rubber (4/Fig. 1). Check the function of the foot switch for motorized compression/decompression (1) and (2) in (Fig. 2). Fig. 2 Foot switch Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 4 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 80: Decompression Button (Only With Separate Control Console)

    Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 4 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 81: Checking The Opcomp

    NOTICE Opcomp has been thoroughly tested by Siemens. If any factors are changed, OPCOMP will no longer be an optimization of com- pression force and image quality according to Siemens’ clinical test.
  • Page 82 15 - 4 Checking the compression device and OPCOMP This page intentionally left blank. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 4 of 4 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 83: Checking The Exposure Blocking

    The film cassette must be exchanged after each exposure. A sec- NOTICE ond exposure before exchanging the casette is not possible (exposure release is blocked). Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 2 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 84 16 - 2 Checking the exposure blocking This page intentionally left blank. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 2 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 85: General

    13 mm. Left edge: 6.5 mm 6.5 mm Right edge: 6.5 mm Back edge: 6.5 mm Front edge: Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 4 Mammomat 1000/3000 Solna, Sweden Rev. 03 11.99 SPS-UD Installation and Start-Up...
  • Page 86: Measuring Procedure

    Write down the values of the centering cross corresponding to the side edges and the front edge of the object table. Release exposure (Mo/Mo). Develop exposed films. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 4 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 87: Evaluation

    . The other sides must be completely exposed Repeat this check with all object tables and both "wings" (base plates). 1. adjust within the given tolerance in point 1 above if necessary Siemens-Elema AB Register 3 SP B7-230.033.07 Page 3 of 4...
  • Page 88 17 - 4 Checking the radiation field limitation This page intentionally left blank. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 4 of 4 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 89: Service Pc And Measuring Instruments

    Main menu: Select Backup <↑> <↓> < ENTER>. Select Copy installation area to floppy <↑> <↓> < ENTER>. Select All <↑> <↓> < ENTER>. Leave the program with <F10>. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 10 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 90: Removing The Measuring Instruments

    (2). Store the strips in the holders provided on both sides of the curtain. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 10 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 91: Mammomat Stand With Separate Rear Side Covers

    (7). Fit inner plastic strip (10) to the front and side covers and fasten with screws (11), eight on either side. Press on the outer plastic strip (12). Siemens-Elema AB Register 3 SP B7-230.033.07 Page 3 of 10...
  • Page 92: Mammomat Stand With Single Rear Cover

    Fit inner plastic strip (7) to the front and side covers and fasten with screws (8), eight on either side. Press on the outer plastic strip (9). Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 4 of 10 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 93: Fitting The Mammomat Cap

    Fitting the cable outlet cover Fig. 4 Cable-outlet cover Fit the cable outlet cover (1) to the stand and fasten with screws (2) and contact washers (3). Siemens-Elema AB Register 3 SP B7-230.033.07 Page 5 of 10 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 94: Mounting The Front Cover Onto The Generator

    (incl mains cable) and any metallic part of the stand and generator which the patient/operator may come in contact with. The protective ground resistance must not exceed 0.2 Ω. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 6 of 10 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 95: Face Shield

    Final procedures 18 - 7 Face shield Fig. 6 Face shield Slide the face shield (1) onto the holder for external diaphragms. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 7 of 10 Mammomat 1000/3000 Solna, Sweden Rev. 03 11.99 SPS-UD...
  • Page 96: Warning Label On Control Panel

    MAM00206 Fig. 7 Warning label on control panel Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 8 of 10 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 97: Affixing The Identification Labels

    11 mm ±1 mm Fig. 8 Affixing the identification labels Returning the LINA card Fill in the LINA card and return it to the address stated in the instructions on the card. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 9 of 10...
  • Page 98 18 - 10 Final procedures This page intentionally left blank. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 10 of 10 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 99: Stativ And Generator With Integrated Radiation Shield And Control Panel

    DIN 9021B, ISO 7093 or equivalent. Tighten the screws to 10 Nm. Make sure that the screws do not come in contact with any con- CAUTION ductive material such as reinforcing iron. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 4...
  • Page 100: Installation With Separate Generator

    Base plate for separate generator Generator secured with wall brackets Fig. 3 Base plate and wall brackets for separate generator Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 4 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 101: Free-Standing Radiation Shield (Option)

    19 - 3 Free-standing radiation shield (option) Fig. 4 Free-standing radiation shield The free-standing radiation shield must be bolted to the floor, see separate instructions enclosed with the radiation shield. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 3 of 4 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 102 19 - 4 Bolting the stand/generator to the floor This page intentionally left blank. This page intentionally left blank. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 4 of 4 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 103: Working With The Service Pc

    <∗> key and then release both keys. <F1> Key <F1> calls up a selective help text. <F10> Key <F10> exits the program. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 1 of 6 Mammomat 1000/3000 Solna, Sweden Rev.
  • Page 104: Connecting The Service Pc

    Save the entered data with <F2>. Page back in the program with <ESC>. The appropriate instructions are shown on the monitor. End the procedure with the service PC with <F10>. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 2 of 6 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 105: Measuring Protocol

    30 kV ______V x ____ s x 40 mA/V = ____ mAs 10 mAs 10 % Date:______________ Sign:______________ Keep measuring protocol in technical documentation. Siemens-Elema AB Register 3 SP B7-230.033.07 Page 3 of 6 Mammomat 1000/3000 Solna, Sweden Rev. 03 11.99...
  • Page 106 ______V x 40 mA/V = ____ mA 125 mA 10 % Date:______________ Sign:____________ Keep measuring protocol in technical documentation. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 4 of 6 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...
  • Page 107 Required values ______V x 5 kV/V = _____kV 30 kV ______V x 40 mA/V = ____ mA 125 mAs 10 % Date:______________ Sign:____________ Siemens-Elema AB Register 3 SP B7-230.033.07 Page 5 of 6 Mammomat 1000/3000 Solna, Sweden Rev. 03 11.99...
  • Page 108 20 - 6 Appendix This page intentionally left blank. Mammomat 1000/3000 Register 3 SP B7-230.033.07 Page 6 of 6 Siemens-Elema AB Installation and Start-Up Rev. 03 11.99 SPS-UD Solna, Sweden...

This manual is also suitable for:

Mammomat 3000