20 - 4
Tube P40 MoW, molybdenum anode 0.15
Calculated values
______V x 5 kV/V = _____kV
______V x ____ s x 40 mA/V = ____ mAs
Tube P40 MoW, tungsten anode 0.3 (MAMMOMAT 3000 only)
40 mm plexiglass
Calculated values
______V x 5 kV/V = _____kV
______V x 40 mA/V = ____ mA
Tube P40 MoW, molybdenum anode 0.3
40 mm plexiglass
Calculated values
______V x 5 kV/V = _____kV
______V x 40 mA/V = ____ mA
Keep measuring protocol in technical documentation.
Mammomat 1000/3000
Installation and Start-Up
Register 3
SP B7-230.033.07
Rev. 03
Required values
30 kV
5 %
10 mAs
10 %
Date:______________
Sign:____________
Required values
30 kV
5 %
158 mA
10 %
Date:______________
Sign:____________
Required values
30 kV
5 %
125 mA
10 %
Date:______________
Sign:____________
Page 4 of 6