Inspection and maintenance
Evaluation:
•
Measure the deviations (X1, X2, Y1, Y2) between the edges of the light field and the
edges of the radiation field on all four sides as shown in the diagram.
•
Calculate the overall difference in the X and Y directions (without regard to the
mathematical sign in front).
•
The length deviation (Σ Y) as well as the width deviation (Σ X) must be less than 1.6 cm
in each case.
NOTE
2.10
Control console
SIE
Operating elements
SIE
Indicators
•
Check the operating elements and indicators on the control console.
- See "Overview of the Operating Elements" in the operator manual.
•
POLYMOBIL OFF and disconnect the power plug.
Siemens AG
Medical Solutions
Y
1
X
2
Y
2
In the case of deviations greater than 1.6 cm, see the
"Coincidence of the Light Field and Radiation Field" chapter
of the service instructions.
X
1
RXR8-120.101.01
Rev. 05
07.05
Film
Light field
Radiation field
Page 11 of 14
CS PS 24
2 - 11
Polymobil 10