Checking the automatic transparency adaption and OPDOSE
(MAMMOMAT 3000 only)
The unit factory-set with the following values, which are to be regarded as recommenda-
tion. Changes can be done with the service PC program (Configuration, Miscellaneous,
Auto limits).
Fig. 5
Installations with filter disc Mo 0.030 / Rh 0.025 /Rh 0.050
Fig. 6
Installations with filter disc Mo 0.030 / Rh 0.025
When making test exposures, follow the instructions on page 11 - 8 for positioning the
plexiglass plates and measuring the density.
NOTICE
Start with 60 mm plexi and compress. Make sure that the display indicates at least
60 mm. Make an exposure and measure the optical density. Repeat this procedure with
50 mm, 40 mm and 20 mm plexi.
Change density with the density setting buttons on the control panel as needed (the panel
programming switch must be ON, see Service PC program Configuration, Miscella-
neous, Panel programming). The optical density for the four different programs shall be
as near as possible to 1.5 O.D. (or the density preferred by the customer). Store with the
memory button.
Mammomat 1000/3000
Installation and Start-Up
Start-up and functional test of the IONTOMAT
Thickness
Th ≥ 60 mm
60 > Th ≥ 45 mm
45 > Th ≥ 30 mm
30 mm > Th
Thickness
Th ≥ 60 mm
60 > Th ≥ 45 mm
45 > Th ≥ 30 mm
30 mm > Th
Both "wings" must be checked, if the stand is equipped with
pivoting object-table arm (Bucky selector).
Register 3
kV
26
27
27
26
kV
28
27
27
26
SP B7-230.033.07
Rev. 03
11.99
Anode/filter
combination
W / Rh
Mo / Rh
Mo / Mo
Mo / Mo
Anode/filter
combination
Mo / Rh
Mo / Rh
Mo / Mo
Mo / Mo
Page 10 of 12
Siemens-Elema AB
SPS-UD
11
Solna, Sweden