5 - 2
Fig. 2
Ferrite sleeves and hose clamp
2.
3.
4.

Note on separate control console and separate generator

If the installation includes a separate control console and separate generator, please refer
to page 5 - 8 for information on how to install the additional connection cable between the
control console and the generator.
Mammomat 1000/3000
Installation and Start-Up
4
Provide each incoming cable with a ferrite sleeve (4), except for mains supply
cables L1 and L2 and the ground wire, which shall be laid together in one and the
same ferrite sleeve. Ferrite sleeves are included in the delivery.
Provide each cable with a hose clamp (3). Hose clamps are included in the
delivery.
Fold the shield braid (1) backwards over the U-bracket (2) on each cable and
secure with the hose clamp (3).
Register 3
3
2
SP B7-230.033.07
Rev. 03
11.99
Cable connections
2
3
1