5 - 12

Connecting the mains supply to the generator (400 V, 2-phase)

Fig. 15 Mains supply to the generator
1.
2.
3.
Mammomat 1000/3000
Installation and Start-Up
Check that there is no wire jumper at F20 (1) and no brass plug at F5 (2).
Fuses must be inserted into fuse holders F20 and F5.
Connect the mains supply cables between stand and generator to fuses F10 and
F20 on the generator.
Wire marked L1 to F10 and L2/N to F20.
Connect the protective ground wire from the stand to the generator (3).
Register 3
F10
SP B7-230.033.07
Rev. 03
11.99
Cable connections
2
1
3
F20