Cable connections

EMC measures

The following measures are required to ensure electromagnetic compatibility of the equip-
ment.

EMC measures at the cable entry

Fitting the hose clamps and ferrite sleeves

Fig. 1 Cables from the stand
1.
Siemens-Elema AB
Solna, Sweden
Lay the cable harness from the stand to the generator and leave approximately
3 cm of the shielding at the U-bracket, see Fig. 1.
Stow away excessive cable lengths as shown in Fig. 2.
Register 3
3cm
SP B7-230.033.07
Rev. 03
11.99