Siemens MOMMOMAT 1000 Installation And Start-Up Instructions Manual page 37

Table of Contents

Advertisement

Cable connections
3.
4.
Ground wire
Fig. 10 Cables from separate control console
5.
Fig. 11 Connecting the cables from the separate control console
Siemens-Elema AB
Solna, Sweden
Shield the control console cable to the bottom plate of the generator as shown in
Fig. 10.
Provide the cables with ferrite sleeves. Cables X5 and X32 can be laid together in
one ferrite sleeve, X30 and the decompression-button cable in another sleeve
(two extra ferrite sleeves are included in the delivery). Plug together the decom-
pression button cables. Shield the decompression button cable to the bottom plate
, see Fig. 10. Lay the decompression button cable in the cable duct, alter-natively
underfloor, to the stand, see Fig. 4 in Chapter 4 "Installing the generator and the
stand".
X5, X32
X30
Connect X5 and X32 to the corresponding sockets on circuit board D700, X30 to
connector T3/X30 at the side of the circuit board chassis. Connect the ground wire
to the protective ground terminal below fuse F10, see Fig. 11.
X32
X5
Register 3
Decompression
button cable
F10
F20
T3/X30
SP B7-230.033.07
Rev. 03
11.99
Protective
ground
Page 9 of 14
Mammomat 1000/3000
SPS-UD
Installation and Start-Up
5 - 9