Adjusting the field light time
The illumination time of the light field can be adjusted according to the customer´s desire
within a certain range.
NOTE
1.
2.
Main menu:
1.
2.
3.
4.
Mammomat 1000/3000
Installation and Start-Up
Shorter illumination time means increased life time of the lamp.
Connect the service PC to the generator and start the program as described in
Chapter 20 "Appendix".
Program according to the following:
Select Configuration <↑> <↓> < ENTER>.
Select Miscellaneous <↑> <↓> <ENTER>.
Select Illumination time <↑> <↓> <ENTER>.
Type in illumination time and save with < F2>.