5 - 8
Installations with separate control console and separate generator
Fig. 8
Connection cable from separate control console
NOTICE
1.
2.
Fig. 9
Shielding the control console cable
Mammomat 1000/3000
Installation and Start-Up
Connection cable from separate control console
Shielding the cable from the separate control console as
described below is an absolute demand to comply with the EMC
regulations.
Lay the connection cable from the separate control console in additional cable
ducts, alternatively underfloor, to the free cable entry of the generator, see Fig. 8.
Lay the connection cable in a free U-bracket. Fold the shield braid over the
U-bracket and secure with a hose clamp as shown in Fig. 9.
Register 3
SP B7-230.033.07
Rev. 03
11.99
Cable connections
Page 8 of 14
Siemens-Elema AB
SPS-UD
5
Solna, Sweden