Reducing The Generator Power; Main Menu - Siemens MOMMOMAT 1000 Installation And Start-Up Instructions Manual

Table of Contents

Advertisement

12 - 2

Reducing the generator power

To achieve full power, the measured resistance must not exceed the following values:
0.25
0.45
0.50
0.60
0.65
0.85
If the above values are exceeded, reduce the generator power as follows:

Main menu:

1.
2.
3.
4.
5.
6.
Mammomat 1000/3000
Installation and Start-Up
at 110 V (1-phase)
at 208 V
at 230 V
at 240 V
at 277 V
at 400 V (2-phase)
Select Configuration <↑> <↓> < ENTER>.
Select Miscellaneous <↑> <↓> < ENTER>.
Select Red. of power <↑> <↓> < ENTER>.
Reduction of power depending on line quality.
Tube is P 40 MoW.
Select nominal line voltage ____ V with cursor keys <↑> <↓> and < ENTER>.
Type in line impedance ____ Ω. Enter the measured line impedance.
Calculated power is ___ kW. Max. possible tube power is displayed.
Save the programmed values with <F2>.
Register 3
Further programming
SP B7-230.033.07
Rev. 03
11.99
Page 2 of 2
Siemens-Elema AB
SPS-UD
12
12
Solna, Sweden

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mammomat 3000

Table of Contents