Reducing the generator power
To achieve full power, the measured resistance must not exceed the following values:
•
0.25
•
0.45
•
0.50
•
0.60
•
0.65
•
0.85
If the above values are exceeded, reduce the generator power as follows:
Main menu:
1.
2.
3.
4.
5.
6.
Mammomat 1000/3000
Installation and Start-Up
at 110 V (1-phase)
at 208 V
at 230 V
at 240 V
at 277 V
at 400 V (2-phase)
Select Configuration <↑> <↓> < ENTER>.
Select Miscellaneous <↑> <↓> < ENTER>.
Select Red. of power <↑> <↓> < ENTER>.
Reduction of power depending on line quality.
Tube is P 40 MoW.
Select nominal line voltage ____ V with cursor keys <↑> <↓> and < ENTER>.
Type in line impedance ____ Ω. Enter the measured line impedance.
Calculated power is ___ kW. Max. possible tube power is displayed.
Save the programmed values with <F2>.