Checking the compression device and OPCOMP
Checking the OPCOMP
1.
2.
Adapting OPCOMP
If the factory-set OPCOMP is not to the customer's satisfaction, an adjustment can be
done with the tautness factors in the "Best compression table", see Service Instructions. A
lower tautness value will increase the compression force and vice versa.
•
Increase/decrease the tautness factor by e.g. 10 the first time. Let the customer use this
setting for at least two weeks, before further changes are made.
NOTICE
Siemens-Elema AB
Solna, Sweden
Decompress.
Set the compression force to 20 kg by using the presetting knob (5/Fig. 1) and
compress.
The compression movement must be interrupted by OPCOMP before 20 kg is
reached, i.e. at 5-15 kg depending on the material (4/Fig. 1) used.
Opcomp has been thoroughly tested by Siemens. If any factors
are changed, OPCOMP will no longer be an optimization of com-
pression force and image quality according to Siemens' clinical
test.
Register 3
SP B7-230.033.07
Rev. 03
11.99
Page 3 of 4
Mammomat 1000/3000
SPS-UD
Installation and Start-Up
15 - 3