Checking The Compression Device - Siemens MOMMOMAT 1000 Installation And Start-Up Instructions Manual

Table of Contents

Advertisement

Checking the compression device and OPCOMP

Checking the compression device

2
1
4
Fig. 1 Swivel-arm system
1.
2.
3.
4.
5.
Fig. 2 Foot switch
Siemens-Elema AB
Solna, Sweden
5
Attach compression plate (1/Fig. 1) if not already done.
Check that the compression plate cannot be removed without pressing the
compression plate release button (2/Fig. 1).
Check the function of the knobs for manual compression/decompression (3/Fig.
1).
Place a piece of soft material on the object table, for example a rolled up towel or a
piece of foam-rubber (4/Fig. 1).
Check the function of the foot switch for motorized compression/decompression
(1) and (2) in (Fig. 2).