Checking The Exposure Blocking - Siemens MOMMOMAT 1000 Installation And Start-Up Instructions Manual

Table of Contents

Advertisement

Checking the exposure blocking

Checking the exposure blocking
Fig. 1 Swivel-arm system
Check that the exposure release is blocked during the following conditions:
No film cassette inserted.
No object table installed.
NOTICE
NOTICE
Siemens-Elema AB
Solna, Sweden
The "casette loaded" switch must be set to "ON" with the service
PC program: Configuration - Miscellaneous - Cassette loaded
check.
The film cassette must be exchanged after each exposure. A sec-
ond exposure before exchanging the casette is not possible
(exposure release is blocked).
Register 3
SP B7-230.033.07
Rev. 03
Page 1 of 2
11.99
SPS-UD
16 - 1
16
16
Mammomat 1000/3000
Installation and Start-Up

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mammomat 3000

Table of Contents