Preliminary work
Removing the stand covers
The MAMMOMAT 3000 stand covers exist in two versions:
•
one with separate rear side covers
•
one with a single rear cover that also covers the rear sides
MAMMOMAT stand with separate rear side covers
Fig. 1
MAMMOMAT with separate rear side covers
CAUTION
•
Elevate swivel-arm system to a comfortable working height.
•
System OFF.
•
Remove plastic strip (1).
•
Remove screws (2) and inner plastic strip (3).
•
Remove screws (4) and front covers (5).
•
Remove screws (6) and rear side covers (7).
Siemens AG
Medical Engineering
7
2
3
1
The metal curtain (see Fig. 1) is very sharp. After removing the
stand front covers (5), mount protective strips onto the edges of
the metal curtain. The strips are located inside the stand.
Register 5
RXB7-230.091.05
Rev. 03
6
4
metal curtain
5
Page 1 of 8
07. 98
TD SD 24
2 - 1
2
2
2
MAMMOMAT 300 / 3000