Bolting the stand/generator to the floor
Stativ and generator with integrated radiation shield and control
panel
Should a permanent fastening to the floor be required due to local regulations (for exam-
ple in earthquake areas) or due to the quality of the floor covering, the stand/generator
can be secured to the floor by means safety dowels and screws. When bolting the stand
to the floor, the fixing screws shall pass through the hollow levelling screws (1). When
bolting the generator with integrated radiation shield and control panel to the floor, the pre-
drilled fixing holes (2) shall be used.
Fig. 1 Base plate for the stand
2
Fig. 2 Base plate for the generator
1.
NOTICE
2.
CAUTION
Siemens-Elema AB
Solna, Sweden
Position the stand /generator and mark out appropriate fixing holes on the floor,
using the base plate as a template.
The base plates shall be secured with at least four screws each
(two on either side).
Drill the holes and insert dowels and screws (use Liebig safety dowel M6 S10/35
and washer DIN 9021B, ISO 7093 or equivalent. Tighten the screws to 10 Nm.
Make sure that the screws do not come in contact with any con-
ductive material such as reinforcing iron.
Register 3
SP B7-230.033.07
Rev. 03
11.99
19
1
2
Page 1 of 4
Mammomat 1000/3000
SPS-UD
Installation and Start-Up
19 - 1