Inspection and maintenance
Fig. 23: Collimator lamp
3.4.2
Checking the illuminance
Required measuring instrument:
- Lux meter
PMF
Illuminance
•
Log the illuminance.
If the light localizer is switched on for illuminating the radiation field, the average illumi-
nance must be at least 180 Lux at a distance of 100 cm. The average illuminance is deter-
mined in that measurements are performed approximately in the center of each of the four
light field quadrants
Fig. 24: Illuminance
3.4.3
Light field/radiation field
Checking the light and radiation field
•
Place the detector (35 cm x 43 cm) on the tabletop.