Light Field/Radiation Field; Evaluation: Light Field To Radiation Field - Siemens MOBILETT XP Digital Installation And Startup

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

Start-up

Light field/radiation field

Checking the light and radiation field
Place the detector (35 cm x 43 cm) on the tabletop.
Set the central beam so it is vertical.
Set a vertical SID of 100 cm; measure to the upper edge of the detector using the tape
measure in the collimator.
Switch on the light localizer and set a light field of approx. 30 cm x 30 cm.
Place the lead ruler (centering cross) centered on the detector.
Measure the light field and make a note of the dimensions
Position a washer as a side marker.
Fig. 16: Centering cross
Create a test patient.
Select an organ program from the "test" range, with approx. 60 kV, 4 mAs.
X
Release an exposure.

Evaluation: light field to radiation field

Using the centering cross, evaluate on screen, on all four sides, the deviations (A, C
and B, D) between the recorded light field and the radiation field edges
Use the zoom function as necessary.
P
Siemens AG
Medical Solutions
The maximum total allowable deviation from the SID is 1.7% (regardless of the
prefix). If the deviation is > 1.7%, the collimator must be adjusted (see the
"Replacement of Parts" instructions).
SPR8-230.814.30.03.02
11.05
(Fig. 16 / p.
MOBILETT XP Digital
Page 37 of 70
CS SD 24
37
0
37).
0
(Fig. 17 / p.
38).