5520A
Operators Manual
STBY
(Standby command) Deactivates the Calibrator output if it is in operate. This is the same
as pressing the Calibrator front panel Y key.
Parameter:
Example:
Disconnect the selected output from the Calibrator front panel terminals.
SYNCOUT
(Synchronization Pulse command) Sends a synchronization pulse out to a slave
Calibrator through the 10 MHZ OUT BNC connector.
Parameter:
Example:
TC_MEAS
(Thermocouple Measure command) Selects the measure thermocouple mode.
Parameters: CEL
Example:
Measure the thermocouple temperature that is attached to the Calibrator TC terminals, in
Celsius.
TC_OFFSET
(Thermocouple Temperature Measurement Offset command) Adds a temperature offset
to thermocouple measurements (±500 °C). This command does not apply to
thermocouple sourcing.
Parameters: <value> CEL
Example:
Add a temperature offset of +10 °C to the thermocouple measurements.
TC_OFFSET?
(Thermocouple Temperature Measurement Offset query) Returns the temperature offset
used for thermocouple measurements (±500 °C).
Responses:
Example:
Returns 10 Celsius when a temperature offset of +10 °C has been added to the
thermocouple measurements.
6-46
x
x
IEEE-488
RS-232
(None)
STBY
x
x
IEEE-488
RS-232
(None)
SYNCOUT
x
x
IEEE-488
RS-232
(Celsius) (optional)
(Fahrenheit) (optional)
FAR
TC_MEAS CEL
x
x
IEEE-488
<value> FAR
TC_O+FFSET +10 CEL
x
IEEE-488
<value> CEL
<value> FAR
TC_OFFSET? returns 1.000E+01,CEL
x
x
Sequential
Overlapped
x
Sequential
x
Sequential
x
RS-232
Sequential
(offset in Celsius) (optional)
(offset in Fahrenheit) (optional)
x
x
RS-232
Sequential
(offset in Celsius) (optional)
(offset in Fahrenheit) (optional)
x
Coupled
x
x
Overlapped
Coupled
x
x
Overlapped
Coupled
x
x
Overlapped
Coupled
x
x
Overlapped
Coupled