Fluke 5520A Operator's Manual page 252

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Operators Manual
Returns Internal, 29.88, Celsius, when the thermocouple reference is internal and at
29.88 °C. (If the temperature return for the internal reference is 0 (0.00E+00), the
Calibrator is not in Operate, and/or the Calibrator is not in a thermocouple mode.)
TC_TYPE
(Thermocouple Type command) Sets the Thermocouple (TC) temperature sensor type.
The TC type is used when the output is set to a temperature value with the OUT
command and the temperature sensor type is set to TC with the TSENS_TYPE
command. When the thermocouple type is changed while simulating a temperature
output, the temperature is changed to 0 °C. Once set, the Calibrator retains the TC type
until power off or reset.
Parameters: B
Example:
Set the thermocouple type for simulating a temperature output to a J-type thermocouple.
6-48
x
x
IEEE-488
RS-232
(B-type thermocouple)
(C-type thermocouple)
C
(E-type thermocouple)
E
(J-type thermocouple)
J
(K-type thermocouple) (default)
K
(N-type thermocouple)
N
(R-type thermocouple)
R
(S-type thermocouple)
S
(T-type thermocouple)
T
(10 µV/°C linear output)
X
(% relative humidity)
Y
(1 mV/°C linear output)
Z
TC_TYPE J
x
x
Sequential
Overlapped
x
Coupled