1-37. AC Current, Square Wave Characteristics (typical)

Range
I < 6 A @ 400 Hz
3 A & 20 A Ranges

1-38. AC Current, Triangle Wave Characteristics (typical)

0.3% of p-p value, from 10% to 90% point
1-34
LCOMP
Risetime
25 µs
off
100 µs
on
Linearity to 400 Hz
Settling Time
40 µs to 1% of final value
200 µs to 1% of final value
< 1% of p-p value, with amplitude > 50% of range
Overshoot
< 10% for < 1 V Compliance
< 10% for < 1 V Compliance
Aberrations