Fluke 5520A Operator's Manual page 109

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

See "Setting AC Voltage Output" above for information on selecting an ac voltage
output in dBm; this procedure assumes an ac voltage output in volts.
Complete the following procedure to set an ac power output. If you make an entry error,
press Gone or more times to clear the display, then reenter the value.
Verify the applied voltage to the UUT does not exceed the
rating of the UUT insulation and the interconnecting wiring.
1. Press R to clear any output from the 5520A.
2. Connect the UUT as described earlier in this chapter under "Connecting the
Calibrator to a UUT." (Adapt the voltage and current connections to suit your
application.)
3. Set the UUT to measure ac power on the desired range.
4. Press the numeric keys and decimal point key to enter the desired voltage output
(maximum six numeric keys). For example, 123.456.
At voltage outputs of 100 volts and above (nominal), you may notice a
slight high-pitched sound. This is normal.
5. Press a multiplier key, if necessary. For example, press c.
6. Press V.
7. The Control Display now shows the amplitude of your voltage entry. For example,
123.456 mV (below).
8. Press the numeric keys and decimal point key to enter the desired current output
(maximum six numeric keys). For example, 234.567.
9. Press a multiplier key, if necessary. For example, press c.
10. Press A.
11. The Control Display now shows the amplitude of your voltage and current entries.
For example, 123.456 mV and 234.567 mA (below).
12. Press the numeric keys and decimal point key to enter the desired frequency output
(maximum five numeric keys). Press a multiplier key, if necessary. For example,
press the kilo multiplier key K. Then press the H key. For example,
1.1234 kHz.
WCaution
Note
Front Panel Operation
Setting the Output
nn084f.eps
nn085f.eps
4
4-29