Fluke 5520A Operator's Manual page 230

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Operators Manual
Parameter:
Example:
Set the instrument into the front panel lockout state. The front panels controls cannot be
used.
LOWS
(Low Potential Output Terminals command) Selects whether or not the Calibrator front
panel NORMAL LO terminal and AUX LO terminal are internally tied together (default)
or are open. This feature is used for ac power, dc power, dual dc volts and dual ac volts
outputs. Once set, the Calibrator retains the LO setting until power off or reset.
Parameter:
Example:
Tie the front panel NORMAL LO and AUX LO terminals together.
LOWS?
(Low Potential Output Terminals query) Returns whether or not the Calibrator front
panel NORMAL LO terminal and AUX LO terminal are internally tied together (default)
or are open.
Response:
Example:
Returns OPEN when the Calibrator front panel NORMAL LO and AUX LO terminals
are not tied together.
MULT
(Multiply command) Multiplies the reference magnitude (as selected with the EDIT
command or default to the primary output). The reference magnitude is the present
reference in either direct mode or in error mode.
Parameter:
6-26
(None)
LOCKOUT
x
x
IEEE-488
RS-232
(disconnect NORMAL LO and AUX LO terminals)
OPEN
(connect NORMAL LO and AUX LO terminals)
TIED
LOWS TIED
x
x
IEEE-488
RS-232
(disconnected NORMAL LO and AUX LO terminals)
OPEN
(connected NORMAL LO and AUX LO terminals)
TIED
LOWS? returns OPEN
x
x
IEEE-488
RS-232
<value>
(multiplier expressed as a floating point number)
x
x
Sequential
Overlapped
x
x
Sequential
Overlapped
x
x
Sequential
Overlapped
x
Coupled
x
Coupled
x
Coupled