Returns the symbolic name of the single or first output, and return the symbolic name of
the second output (0 if there is no second output).
RANGELCK
(Range Lock command) Locks in the present range, or selects auto ranging for dc
voltage and dc current single outputs. The range automatically unlocks if the output
function changes, for example from dc volts to dc current. When RANGELCK is on, this
is equivalent to the softkey range lock showing locked. When RANGELCK is off, this is
equivalent to the softkey range lock showing auto.
Parameter:
Example:
Set the range lock off to allow autoranging for dc volts or dc current.
RANGELCK?
(Range Lock query) Returns whether or not the preset dc volts or dc current single
output range is locked.
Response:
Example:
Returns OFF when the range for dc volts or dc current is not locked (autoranging
enabled).
6-36
DC330MV,0
DC33MA_A,0
AC3_3V,0
AC330MA_A,0
R110OHM,0
C1_1UF,0
TCSRC,0
RTD_110,0
DC3_3V_P,DC3A_AS
AC330V_P,AC20A_2S
DC330MV_P,DC3_3V_S
AC330V_P,AC3_3V_S
x
x
IEEE-488
RS-232
(Locks the dc volts or dc current range)
ON
(Unlocks the dc volts or dc current range for autoranging)
OFF
RANGELCK OFF
x
x
IEEE-488
RS-232
(range is locked and autoranging is not allowed)
ON
(range is not locked and autoranging is allowed)
OFF
RANGELCK? returns OFF
(dc volts 330 mV range)
(dc current 33 mA range)
(ac volts 3.3 V range)
(ac current 330 mA range)
(ohms 110 Ω range)
(capacitance 1.1 µF range)
(temperature thermocouple source)
(temperature RTD 110 Ω range)
(dc power 3.3 V, 3 A ranges)
(ac power 330 V, 20 A ranges)
(dual dc volts 330 mV, 3.3 V ranges)
(dual ac volts 330 V, 3.3 V ranges)
x
x
Sequential
Overlapped
x
x
Sequential
Overlapped
x
Coupled
x
Coupled