5520A
Operators Manual
9. Verify that the harmonic amplitude and phase angle readings displayed by the Tester
10. Repeat the previous three steps using the settings and limits in Table 4-4.
Amplitude
11. Press Y to remove the voltage from the Tester.
4-70
are within the minimum and maximum limits listed in Table 4-4. (Note: The Tester
will read a positive phase when the 5520A output is a negative phase because, on the
5520A, the polarity of the phase is always relative to the NORMAL channel output.)
Table 4-4. Harmonics Performance for Volts, Harmonics Scree
5520A
Normal Output
Harmonic
(V)
No.
7.00
1
7.00
3
7.00
9
7.00
13
7.00
21
7.00
31
Fluke
Tester
Harmonic
Phase
cursor
(deg.)
No.
MIN
-10
1
-20
3
-30
9
-40
13
-50
21
-60
31
Performance Limits
Amplitude
Phase
MAX
MIN
6.7
7.3
8
6.7
7.3
14
6.7
7.3
21
6.7
7.3
29
6.5
7.5
35
6.2
7.8
40
MAX
12
26
39
51
65
80