Fluke 5520A Operator's Manual page 234

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Operators Manual
Examples:
Each example shows a value and unit, e.g., −15.2 V. If a value is entered without a unit,
the value of the existing output is changed, when logically allowed.
OUT?
(Output query) Returns the output amplitudes and frequency of the Calibrator.
Multipliers (e.g., K or M) are not used in the response.
Parameters: V
Response:
Examples:
6-30
OUT 15.2 V
OUT 20 DBM
OUT 10 V, 60 Hz
OUT 10 DBM, 50 HZ
OUT 1.2 MA
OUT 1 A, 400 HZ
OUT 1 KOHM
OUT 1 UF
OUT 100 CEL
OUT −32 FAR
OUT 60 HZ
OUT 10 V, 1 A
OUT 1 V, 1 A, 60 HZ
OUT 1 V, 2 V
OUT 10 MV, 20 MV, 60 HZ (dual volts; .01 V, .02 V @ 60 Hz)
x
x
IEEE-488
RS-232
(optional for ac voltage and TC modes)
(optional for ac voltage modes)
DBM
(optional for RTD and TC modes, Celsius)
CEL
(optional for RTD and TC modes, Fahrenheit)
FAR
(optional for RTD modes, ohms)
OHM
<primary amplitude value>,<primary units>,
<secondary amplitude value>,<secondary units>,
<fundamental frequency value>
OUT? returns −1.520000E+01,V,0E+00,0,0.00E+00
OUT? returns 1.88300E−01,A,0E+00,0,4.420E+02
OUT? returns 1.23000E+00,V,2.34000E+00,V,6.000E+01
OUT? returns 1.92400E+06,OHM,0E+00,0,0.00E+00
OUT? returns 1.52000E+01,V,1.88300E−01,A,4.420E+02
OUT? DBM returns 2.586E+01,DBM,0E+00,A,4.420E+02
OUT? returns 1.0430E+02,CEL,0E+00,0,0.00E+00
OUT? FAR returns 2.19740000E+02,FAR,0E+00,0,0.00E+00
OUT? V returns 4.2740E−03,V,0E+00,0,0.00E+00
OUT? OHM returns 1.40135E+02,OHM,0E+00,0,0.00E+00
(volts; 15.2 V @ same frequency)
(volts; 20 dBm @ same frequency)
(volts ac; 10 V @ 60 Hz)
(volts ac; 10 dBm @ 50 Hz)
(current; 1.2 mA @ same frequency)
(current ac; 1 A @ 400 Hz)
(ohms; 1 kΩ)
(capacitance; 1µF)
(temperature; 100 °C)
(temperature; −32°F)
(frequency update; 60 Hz)
(power; 10 watts @ same frequency)
(power ac; 1 watts @ 60 Hz)
(dual volts; 1 V, 2 V @ same freq.)
x
x
Sequential
Overlapped
x
Coupled

Advertisement

Table of Contents
loading

Table of Contents