Pulse Response Calibration Procedure For An Oscilloscope - Fluke 5520A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

8-100. Pulse Response Calibration Procedure for an Oscilloscope

This sample procedure shows how to check the oscilloscope's pulse response. Before
you check your oscilloscope, see your oscilloscope's manual for the recommended
calibration settings.
Before you start this procedure, verify that you are running the oscilloscope option in
Edge mode. If you are, the Control Display shows the following menu.
Perform the following sample procedure to calibrate the pulse response.
1. Connect the 5520A to Channel 1 on the oscilloscope. Select 50Ω impedance or use a
50Ω termination directly at the oscilloscope input. Verify that the O key is lit,
indicating that the signal is connected.
2. Alter the voltage setting for the signal so it matches the amplitude value
recommended by your oscilloscope manufacturer for calibrating the edge response.
The default setting is 25 mV @ 1 MHz.
For example, on a Fluke PM3392A oscilloscope, start with a signal of 1 V @ 1
MHz.
3. Adjust the scale on your oscilloscope to achieve a good picture of the edge. For
example, on a Fluke PM3392A oscilloscope with a 1 V @ 1 MHz signal, use
200 mV/div.
4. Adjust the time base on your oscilloscope to the fastest position available (20.0 or
50.0 ns/div).
5. Verify that your oscilloscope exhibits the proper rise time and pulse aberration
characteristics.
6. Remove the input signal by pressing Y.
Calibrating the Pulse and Frequency Response on an Oscilloscope
Pulse aberrations
5520A-SC300 Option
gl028i.eps
gl007i.eps
8
8-73

Advertisement

Table of Contents
loading

Table of Contents