Fluke 5520A Operator's Manual page 368

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Operators Manual
SCOPE
Programs the oscilloscope calibration option hardware, if installed. The instrument
settings are determined by this command's parameter. Once in SCOPE mode, use the
OUT command to program new output.
OPER, STBY, *OPC, *OPC?, and *WAI all operate as described in Chapter 6. The state
of the oscilloscope's output while in SCOPE mode is reflected by the bit in the ISR that
is assigned to SETTLED.
The FUNC? query returns SDCV, SACV, LEVSINE, MARKER, EDGE, and WAVEGEN for
the corresponding oscilloscope modes.
Parameters: OFF
Example:
8-84
x
x
IEEE-488
RS-232
VOLT
EDGE
LEVSINE
MARKER
WAVEGEN
SCOPE VOLT;
SCOPE VOLT;
SCOPE EDGE;
SCOPE LEVSINE; OUT 1 V, 20 kHz
SCOPE MARKER;
SCOPE WAVEGEN; OUT 1 V, 1 kHz
x
x
Sequential
Overlapped
Turns the oscilloscope hardware off. Programs 0 V,
0 Hz, output at the NORMAL terminals, standby.
Oscilloscope ac and dc voltage mode.
Programs 20 mV p-p, 1 kHz, output at the
SCOPE BNC, output impedance 1 MΩ, standby if
from OFF or previously in standby.
Oscilloscope Edge mode. Programs 25 mV peak-to-
peak, 1 MHz, output at the SCOPE BNC, standby if
from OFF or previously in standby.
Oscilloscope-leveled sine mode. Programs 30 mV
p-p, 50 kHz, output at the SCOPE BNC,
standby if from OFF or previously in standby.
Oscilloscope Marker mode. Programs the period to
1 ms, output at the SCOPE BNC, standby if from
OFF or previously in standby.
Oscilloscope Wavegen mode. Programs 20 mV
p-p, square wave, 1 kHz, no offset, output
impedance 1 MΩ, standby if from OFF or
previously in standby.
OUT -2 V, 0 Hz
OUT 4 V, 1 kHz
OUT 0.5 V, 5 kHz (Edge, 0.5 V p-p,
OUT 2 MS
x
Coupled
(dc voltage, -2 V)
(ac voltage, 4 V peak-to-
peak, 1 kHz.)
5 kHz.)
(Leveled sine wave, 2 V
p-p, 20 kHz.)
(Marker, period of 2 ms.)
(Wave Generator, 1 V
p-p, 1 kHz.)