Table 7-13. Verification Tests for Phase Accuracy, V and I
Range,
Normal
Output,
Output
Normal
329.999 mV
30.000 mV
329.999 mV
30.000 mV
329.999 mV
30.000 mV
329.999 mV
200.000 mV
329.999 mV
50.000 mV
329.999 mV
50.000 mV
329.999 mV
30.000 mV
329.999 mV
200.000 mV
329.999 mV
200.000 mV
329.999 mV
200.000 mV
32.9999 V
3.3000 V
32.9999 V
3.3000 V
32.9999 V
3.3000 V
32.9999 V
3.3000 V
32.9999 V
3.3000 V
32.9999 V
3.3000 V
32.9999 V
3.3000 V
32.9999 V
3.3000 V
329.999 V
33.000 V
329.999 V
33.000 V
329.999 V
33.000 V
329.999 V
33.000 V
329.999 V
33.000 V
329.999 V
33.000 V
329.999 V
33.000 V
329.999 V
33.000 V
Freq-
Range, AUX
uency
Output
Output, AUX
65 Hz
329.99 mA
300.00 mA
1 kHz
329.99 mA
300.00 mA
30 kHz
329.99 mA
300.00 mA
65 Hz
2.99999 A
2.00000 A
65 Hz
20.5000 A
5.0000 A
400 Hz
20.5000 A
5.0000 A
65 Hz
329.99 mA
300.00 mA
65 Hz
2.99999 A
2.00000 A
65 Hz
20.5000 A
20.0000 A
400 Hz
20.5000 A
20.0000 A
65 Hz
329.99 mA
300.00 mA
65 Hz
2.99999 A
2.00000 A
65 Hz
20.5000 A
5.0000 A
400 Hz
20.5000 A
5.0000 A
65 Hz
329.99 mA
300.00 mA
65 Hz
2.99999 A
2.00000 A
65 Hz
20.5000 A
20.0000 A
400 Hz
20.5000 A
20.0000 A
65 Hz
329.99 mA
300.00 mA
65 Hz
2.99999 A
2.00000 A
65 Hz
20.5000 A
5.0000 A
400 Hz
20.5000 A
5.0000 A
65 Hz
329.99 mA
300.00 mA
65 Hz
2.99999 A
2.00000 A
65 Hz
20.5000 A
20.0000 A
400 Hz
20.5000 A
20.0000 A
Maintenance
Performance Tests
Lower
Upper
Phase
Limit
Limit
°
°
0
-0.10
0.10
0
-0.50
0.50
0
-10.00
10.00
0
-0.10
0.10
0
-0.10
0.10
0
-0.25
0.25
60
59.90
60.10
60
59.90
60.10
60
59.90
60.10
60
59.75
60.25
0
-0.10
0.10
0
-0.10
0.10
0
-0.10
0.10
0
-0.25
0.25
90
89.90
90.10
90
89.90
90.10
90
89.90
90.10
90
89.75
90.25
0
-0.10
0.10
0
-0.10
0.100
0
-0.10
0.10
0
-0.25
0.25
90
89.90
90.10
90
89.90
90.10
90
89.90
90.10
90
89.75
90.25
7
°
7-21