Fluke 5520A Operator's Manual page 116

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Operators Manual
4-36
V @ NOR (Voltage at NORMAL Terminals) V @ AUX (Voltage at AUX
Terminals) This is an information-only softkey position and does not have an
associated function. It shows the output function is dual ac voltage.
WAVE MENUS (Waveform Menus) Opens submenus for selecting the type of
harmonic, waveform, front panel LO terminal condition, and phase.
HARMONIC MENUS (Harmonic Frequency Menus) Opens submenus for
selecting harmonic outputs. See "Setting Harmonics" later in this chapter for
more information.
WAVE (Normal Waveform) Selects the waveform for the voltage at the front
panel NORMAL terminals. See "Waveform Types" later in this chapter for more
information.
AUXWAVE (Auxiliary Waveform) Selects the waveform for the voltage at the
front panel AUX terminals. See "Waveform Types" later in this chapter for more
information.
"LO"s (Low Potential Output Terminals) The front panel NORMAL LO and
AUX LO terminals must be tied together either at the UUT or at the 5520A.
When the front panel NORMAL LO and AUX LO terminals are tied at the UUT,
select "open" with the "LO"s softkey. If the NORMAL LO and AUX LO
terminals are not tied at the UUT, select "tied" with the "LO"s softkey. The
default is tied.
φ & REF MENUS (Phase Difference and 10 MHz reference source.) Selects the
phase difference between the NORMAL and AUX outputs, selects internal or
nn095f.eps