Calibrating The Time Base Of An Oscilloscope; The Time Marker Function - Fluke 5520A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Operators Manual
6. Remove the input signal by pressing Y.
7. Repeat this procedure for the remaining channels on your oscilloscope.

8-39. Calibrating the Time Base of an Oscilloscope

The horizontal deflection (time base) of an oscilloscope is calibrated using a method
similar to the vertical gain calibration. A time marker signal is generated from the
Calibrator and the signal's peaks are matched to the graticule line divisions on the
oscilloscope.

8-40. The Time Marker Function

The Time MARKER function, which is available through the MARKER menu, lets you
calibrate the timing response of your oscilloscope. To access the MARKER menu, press
the softkey under MODE until "marker" appears.
You can press the MODE softkey to cycle through the functions in the order shown, or
you can press P to return directly to the SCOPE menu.
Each option in the MARKER menu is described below:
8-26
Output at SCOPE
terminal (50Ω).
f
f
OUTPUT @ SCOPE terminal (50Ω) Indicates the location of the signal output. If
the signal does not appear on the oscilloscope, press O. To disconnect the signal,
press Y.
SHAPE
TRIG
spike
off
f
f
sine
off
spike
/1
square
/10
sq20%
/100
gl010i.eps
MODE
marker
f
volt
edge
levsine
marker
wavegen
video
pulse
meas Z
overld