The entered temperature will be cleared to 0 ° C (32 ° F) if you change
between tc and rtd, or change the type of thermocouple (except for a type B
thermocouple, which clears to 600 ° C). If this should occur, select
OUTPUT tc, the desired thermocouple TYPE, and then reenter the
temperature.
Out@TC terminal (Output at the front panel TC terminals) Displays the actual dc
voltage at the front panel TC terminals. This is a display only, not a softkey function.
TC MENUS (Thermocouple Menu) Shows submenus for thermocouple outputs.
UNITS (Temperature Units) Selects °C or °F as the temperature unit.
REF SRC (Reference Source) Selects intrnl (Internal) or extrnl (External)
temperature reference source. Select intrnl when the selected thermocouple has
Note
Front Panel Operation
Setting the Output
nn104f.eps
4
4-41