Fluke 5520A Operator's Manual page 369

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

x
SCOPE?
IEEE-488
Returns the oscilloscope's current mode of operation. Returns OFF if the oscilloscope is
off.
Parameter:
(None)
Response:
<character>
x
TRIG
IEEE-488
Programs the oscilloscope's trigger output BNC.
Parameters: OFF
DIV1
DIV10
DIV100
x
TRIG?
IEEE-488
Returns the output setting of the oscilloscope's trigger.
Parameters: (None)
Response:
<character>
x
OUT_IMP
IEEE-488
Programs the oscilloscope's output impedance.
Parameters: Z50
Z1M
x
OUT_IMP?
IEEE-488
Returns the impedance setting of the oscilloscope's output.
Parameters: (None)
Response:
<character>
x
x
RS-232
Sequential
(Returns OFF, VOLT, EDGE, LEVSINE, MARKER, or
WAVEGEN.)
x
x
RS-232
Sequential
(Turns the trigger output off.)
(Turns the trigger output on. Frequency is the same as the
signal at SCOPE output.)
(Turns the trigger output on. Frequency is 1/10 of the
signal at SCOPE output.)
(Turns the trigger output on. Frequency is 1/100 of the
signal at SCOPE output.)
x
x
RS-232
Sequential
(Returns OFF, DIV1, DIV10, or DIV100.)
x
x
RS-232
Sequential
(Programs the oscilloscope's output impedance to
50Ω.)
(Programs the oscilloscope's output impedance to
1 MΩ.)
x
x
RS-232
Sequential
(Returns Z50 or Z1M.)
5520A-SC300 Option
Summary of Commands and Queries
x
x
Overlapped
x
x
Overlapped
Coupled
x
x
Overlapped
Coupled
x
x
Overlapped
x
Overlapped
8
Coupled
Coupled
x
Coupled
8-85